Why molecules absorb in the UV and visible range
The ultraviolet band from roughly 200 to 400 nm and the visible band from about 400 to 780 nm together cover the energies that move electrons between orbitals. A molecule absorbs a photon only when the photon energy matches the gap between two levels. Groups such as double bonds, aromatic rings and carbonyls, which are called chromophores, are the parts of a molecule responsible for this absorption. Each substance has a wavelength where absorption peaks, and measurements are usually taken there because sensitivity is highest.
How the instrument is laid out
In a double beam instrument light leaves the source, passes through a monochromator that selects one wavelength, and is split so that one part crosses the sample and the other crosses a reference. A detector then compares the two beams. Because sample and reference are read together, fluctuations of the lamp and the solvent largely cancel out. Quartz cuvettes are used in the ultraviolet range, because ordinary glass absorbs that light and would distort the reading.
Absorbance, transmittance and the Beer-Lambert law
Transmittance T is the fraction of light that passes through a solution, and absorbance A is defined as minus the logarithm of T. A higher absorbance therefore means less light reaches the detector. The Beer-Lambert law states that A equals molar absorptivity times path length times concentration. Doubling the cuvette length has the same effect as doubling the concentration, which is why the path length is always stated next to the result.
Building a calibration curve
A calibration curve is built from at least five standards of known concentration. Each standard is measured at the same wavelength and the absorbance values are plotted against concentration. A straight line fitted through the points gives a relation of the form A = intercept + slope × c. The intercept should be included in the calculation: ignoring it shifts the answer for every unknown, even when the slope is correct.
A worked example with a new calibration line
Take the calibration A = 0.015 + 0.080 × c, where c is in mg/L. An unknown sample reads 0.335. Subtract the intercept: 0.335 − 0.015 = 0.320. Divide by the slope: 0.320 / 0.080 = 4.0 mg/L. If the intercept were ignored, the answer would be 0.335 / 0.080 = 4.19 mg/L, a clearly larger error that comes from a single skipped step.
Dilution and the original concentration
When a sample is too concentrated to fall within the calibration range, it is diluted first. If 5.0 mL of sample is made up to 50.0 mL, the dilution factor is 10. The reading obtained from the diluted solution is multiplied by that factor to recover the original concentration. Forgetting this step is one of the most common reasons a correct measurement produces a wrong final answer.
Frequently asked questions
Which cuvette material should be used in the ultraviolet range?
Quartz is used in the ultraviolet range because ordinary glass absorbs much of that light and distorts the reading. In the visible range, glass cuvettes are usually adequate.
Why does the calibration curve need an intercept?
Because the blank and the matrix can add a constant absorbance even when the concentration is zero. Leaving the intercept out shifts every calculated concentration, so the fitted line should always include it.
What should be done when the reading is above the calibration range?
Dilute the sample by a known factor, measure it again inside the calibrated range, and multiply the result by the dilution factor. Extrapolating the line far beyond the highest standard is not reliable.