One goal, two signals
ECG and EMG are the two biosignal systems most often met in medical device service. Both aim to pick up a very small electrical potential difference from the body without disturbing patient safety, then remove noise, convert the signal to digital form and display it at the correct scale. What separates them is the source of the signal and therefore its amplitude and frequency range.
The ECG measurement chain
In ECG the signal comes from depolarisation and repolarisation of heart muscle, typically 0.5 to 5 mV on the body surface. The chain runs through a silver chloride gel electrode, a shielded lead, an instrumentation amplifier, analogue filters, an analogue-to-digital converter and display software. The first stage needs very high input impedance and very low noise, because later stages amplify noise along with the signal.
The EMG measurement chain
EMG records motor unit action potentials from skeletal muscle. Surface EMG uses electrodes on the skin and yields microvolt-range signals, typically 50 microvolts to 5 millivolts, with a band of about 20 to 500 Hz. Needle EMG places a sterile electrode directly in the muscle, which gives a more local signal, but it is an invasive procedure that must be performed by trained staff.
Comparing the two signals
A useful way to remember the difference is to compare amplitude, bandwidth and sampling. Diagnostic ECG covers 0.05 to 150 Hz and is typically sampled at 500 Hz or more. EMG reaches higher frequencies and therefore needs a higher sampling rate, at least 1000 Hz in the surface case. Electrode type and filter settings must match the signal, not the other way round.
Gain and resistor calculation
An ECG preamplifier may need to lift a 1 mV input to 100 mV, so the required gain is 100. In a non-inverting stage, gain equals 1 plus Rf over R1; with R1 at 1 kΩ, Rf becomes 99 kΩ. Common-mode rejection matters too: a CMRR of 80 dB corresponds to about 10,000 times suppression of common-mode noise.
Patient safety and device class
Electrodes that reach the heart are classed as type CF applied parts, where the normal patient leakage current limit is 10 µA. The IEC 60601 series, part 2-25, covers ECG equipment requirements. Calibration uses a 1 mV pulse that should produce a 10 mm deflection, which means 10 mm per mV. Checking this scale is a routine acceptance test.
Frequently asked questions
How many millimetres of deflection should a 1 mV calibration pulse produce on an ECG?
The standard sensitivity is 10 mm per mV, so a 1 mV calibration pulse should give exactly 10 mm of deflection on the recorder.
Which applied part type is used for electrodes that reach the heart?
Electrodes that reach the heart use type CF, where the normal patient leakage current limit is 10 µA.
An amplifier has a CMRR of 80 dB. By what factor is common-mode noise reduced?
From CMRR in dB = 20 × log10(K), 80 = 20 × log10(K) gives K = 10 to the power 4, that is a reduction of 10,000 times.